A dexterous system for laryngeal surgery

Nabil Simaan, Russell H. Taylor, Paul W. Flint

2004 · 314 citations · 22 references

Concepts

TL;DR

The paper presents a design overview of a novel high‑degree‑of‑freedom system for minimally invasive throat surgery, outlining kinematic analysis and proposing actuation redundancy to enable scalability and reduce backbone buckling. The system comprises a 34‑DoF unit with three snake‑like distal dexterity modules, each a multi‑backbone snakelike mechanism equipped with a detachable milli‑parallel manipulator for interchangeable tools, allowing remote operation of 2–3 tools with high tip dexterity through the patient’s mouth. An early experiment with a prototype snake‑like unit demonstrates the feasibility of the design.

Abstract

This work presents a design overview of a novel high DoF (degrees-of-freedom) system being developed for minimally invasive surgery of the throat. The system is designed to allow remote operation of 2-3 tools with high tip dexterity to enable suturing and soft-tissue manipulation while using the patient's mouth as the only entry port. The slave is a 34 DoF unit equipped with three snake-like distal dexterity units for surgical tool manipulation. Each of these units is a multi-backbone snakelike mechanism equipped with a detachable milli parallel manipulator allowing interchangeable tools to be used. The paper presents the outline of the kinematic analysis of the snake-like units and proposes one possible actuation redundancy resolution to allow further downsize scalability while reducing the risk of buckling of the primary backbone of the snake-like units. Finally, The work presents a first early experiment with a prototype of the snake-like unit.

References

22