2015 · 32 citations · 25 references
EngineeringNeurosurgerySurgeryDiagnostic ImagingMedical RoboticsRoboticsComputed TomographyImage-guided RoboticsInterventional RoboticsRobot-assisted SurgeryStroke SurgerySurgical RoboticsDeformationHemorrhagic Stroke SurgeryImage-guided Hematoma RemovalImage-guided Robotic SurgeryIntracerebral Hemorrhage EvacuationRobotic NeurosurgeryRobotic-assisted Hematoma EvacuationSurgical Planning
Several robotic systems have been proposed for removing blood from the brain in patients who have undergone a hemorrhagic stroke. In this paper we explore the use of imagebased feedback to address tissue deformation when aspirating a hemorrhage in a phantom model. This is the first time intraoperative image feedback has been used with a concentric tube robot in this application. We describe a layer by layer approach to motion planning. Computed tomography (CT) images are collected periodically during hemorrhage removal. After each CT scan, the robot's tip path is re-planned to account for the tissue deformation that has occurred since the previous scan. We compare open loop hemorrhage removal to our sequential imaging-replanning approach, illustrating that the latter has the potential to enhance the safety and efficacy of the procedure.
25
Guidelines for the Management of Spontaneous Intracerebral Hemorrhage in Adults
Joseph P. Broderick, Sander Connolly, Edward Feldmann et al. · Stroke · 2007 · 1.1K citations · Full text
Design and Control of Concentric-Tube Robots
Pierre E. Dupont, Jesse Lock, Brandon Itkowitz et al. · IEEE Transactions on Robotics · 2010 · 737 citations · Full text
Engineering, Mechanical Engineering, Robotics Engineering +16