Fast‐Response Variable‐Stiffness Magnetic Catheters for Minimally Invasive Surgery

Yegor Piskarev, Yi Sun, Matteo Righi, Quentin Boehler, Christophe Chautems, Cedric Fischer, Bradley J. Nelson, Jun Shintake, Dario Floreano

Advanced Science · 2024 · 51 citations · 47 references

DOIFull text

Open access

Concepts

Abstract

In minimally invasive surgery, such as cardiac ablation, magnetically steered catheters made of variable-stiffness materials can enable higher dexterity and higher force application to human tissue. However, the long transition time between soft and rigid states leads to a significant increase in procedure duration. Here, a fast-response, multisegmented catheter is described for minimally invasive surgery made of variable-stiffness thread (FRVST) that encapsulates a helical cooling channel. The rapid stiffness change in the FRVST, composed of a nontoxic shape memory polymer, is achieved by an active cooling system that pumps water through the helical channel. The FRVST displays a 66 times stiffness change and a 26 times transition enhancement compare with the noncooled version. The catheter allows for selective bending of each segment up to 127° in air and up to 76° in water under an 80 mT external magnetic field. The inner working channel can be used for cooling an ablation tip during a procedure and for information exchange via the deployment of wires or surgical tools.