Amperometric Enzyme Carbon Paste-Based Biosensor for Quantification of Hydroquinone and Polyphenolic Antioxidant Capacity

Milan Sýs, Bruna Pekec, Kurt Kalcher, Karel Vytřas

International Journal of Electrochemical Science · 2013 · 25 citations · 19 references

DOIFull text

Open access

Concepts

Abstract

An amperometric biosensor for the determination of hydroquinone and other polyphenolic antioxidants is presented, based on a carbon paste electrode on which tyrosinase is immobilized in a Nafion® film. In a 0.1 M phosphate buffer (pH 7.0 and 25 º 1 °C), the antioxidant capacity can be expressed as the concentration equivalent of p-hydroquinone (HQ), according to an amperometric response of polyphenolic antioxidants with the carbon paste sensor at –0.24 V vs. Ag/AgCl/3 M KCl. In the range of low HQ concentrations (up to 120 μmol L-1), the corresponding calibration curve follows the equation I [μA] = – 0.0426 – 0.0113 c(HQ) [μmol L-1] with a correlation coefficient R = 0.999. The limit of detection (LOD for S/N = 3) was estimated as 1.6 μmol L-1, the repeatability was 1.2 % RSD (c = 90 μM, n = 5 measurements). The biosensor can be applied to determine the phenolic antioxidant capacity in plants and food samples in the absence of higher concentrations of ascorbic acid.

References

19