IEEE Transactions on Nuclear Science · 2010 · 20 citations · 20 references
EngineeringElectrical EngineeringMetrologyRadiation MeasurementSensorsSensor CalibrationTemperature MeasurementTemperature SensorsBias Temperature CompensationDosimeter AccuracyDosimeter Response CharacterizationDosimetersRadiation DosimetersSilicon DosimetryTemperature CompensationThreshold Voltage
<para xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> Temperature dependence of MOS dosimeters response used under the Bias Controlled Cycled Measurement technique is investigated. The use of the biasing technique allows the compensation of temperature-induced changes in the response of the sensors, and reduces at least ten times the dose measurement error caused by undesired threshold voltage shifts. </para>
20
Electronic dosimetry in radiation therapy
Anatoly Rosenfeld · Radiation Measurements · 2006 · 122 citations