IEEE Transactions on Medical Robotics and Bionics · 2023 · 11 citations · 40 references
EngineeringBiomedical EngineeringRobotics EngineeringMedical RoboticsRobot ControlRoboticsInterventional RoboticsSoft Robot ControlContinuum RoboticsDeflection ControlDynamic ResponseModel Reference Adaptive ControlNitinolPid TuningTendon-driven ActuationTendon-driven Actuators
Steerable needles are desired for many applications within the medical field. One such application is high-dose rate (HDR) brachytherapy. In this paper, an adaptive control system is proposed to provide control of the planar bending angle or deflection distance of a tendon-driven continuum joint, which can be inserted as the stylet in a commercial hollow needle for HDR brachytherapy. Proportional–integral–derivative (PID) controllers are a common method of control. However, significant tuning may be required to achieve the desired dynamic response, and the tuned gains may not achieve the desired response in a different environment or if the system model varies. Steerable needles must navigate through tissues of differing viscoelastic properties and the use of nitinol for the joint and tendons of a steerable needle leads to change in the system parameters due to change in the phase of the nitinol material. In this article, we propose the real-time adaptation of PID gains using model reference adaptive control techniques to control the planar bending angle or deflection distance of a steerable stylet and needle system.
40
Elastic modulus measurements of human liver and correlation with pathology
Wen‐Chun Yeh, Pai‐Chi Li, Yung‐Ming Jeng et al. · Ultrasound in Medicine & Biology · 2002 · 459 citations
Medicine, Biophysics, Pathology +14
Amar U. Kishan, Ryan Cook, Jay P. Ciezki et al. · JAMA · 2018 · 314 citations · Full text