Journal of Biomechanical Engineering · 2006 · 155 citations · 10 references
EngineeringMechanical EngineeringRobotics EngineeringActuatorsOrthoticsRoboticsAnkle BiomechanicsAnkle TendonsLinear ActuatorsLower Limb OrthosesOrthotic BiomechanicsOrthotic DevicesEnergy-efficient Gait GenerationEnergy-efficient LocomotionGaitGait Assistance DevicesGearbox SystemsLinear Actuator EfficiencyRobotic OrthosesTendon Stiffness
A robotic tendon is a spring based, linear actuator in which the stiffness of the spring is crucial for its successful use in a lightweight, energy efficient, powered ankle orthosis. Like its human analog, the robotic tendon uses its inherent elastic nature to reduce both peak power and energy requirements for its motor. In the ideal example, peak power required of the motor for ankle gait is reduced from 250 W to just 77 W. In addition, ideal energy requirements are reduced from nearly 36 J to just 21 J. Using this approach, an initial prototype has provided 100% of the power and energy necessary for ankle gait in a compact 0.95 kg package, seven times less than an equivalent motor/gearbox system.
10
Marc H. Raibert, Ernest R. Tello · IEEE Expert · 1986 · 2.7K citations
Gait Analysis — An Introduction
David Condie · International Journal of Rehabilitation Research · 1992 · 658 citations
The RoboKnee: an exoskeleton for enhancing strength and endurance during walking
Jerry Pratt, Ben Krupp, Chris Morse et al. · 2004 · 563 citations
Engineering, Mechatronics Engineering, Robotics Engineering +17
Series elastic actuator development for a biomimetic walking robot
David Robinson, Jerry Pratt, Daniel Paluska et al. · 1999 · 415 citations
Engineering, Mechanical Engineering, Robotics Engineering +17