Journal of Medical Robotics Research · 2018 · 46 citations · 40 references
EngineeringRobotics EngineeringSurgeryMedical RoboticsRobot ControlRoboticsImage-guided BiopsyInterventional RoboticsProstate BiopsyProstate SurgeryRobot-assisted SurgerySurgical RoboticsMagnetic Resonance Imaging - Targeted BiopsyMri-targeted BiopsyNeedle Placement AccuracyRobotic Biopsy GuidanceRobotic Biopsy SystemsSurgical Navigation SystemsTargeted Prostate BiopsyTransperineal Biopsy
This paper presents the development, preclinical evaluation, and preliminary clinical study of a robotic system for targeted transperineal prostate biopsy under direct interventional magnetic resonance imaging (MRI) guidance. The clinically integrated robotic system is developed based on a modular design approach, comprised of surgical navigation application, robot control software, MRI robot controller hardware, and robotic needle placement manipulator. The system provides enabling technologies for MRI-guided procedures. It can be easily transported and setup for supporting the clinical workflow of interventional procedures, and the system is readily extensible and reconfigurable to other clinical applications. Preclinical evaluation of the system is performed with phantom studies in a 3 Tesla MRI scanner, rehearsing the proposed clinical workflow, and demonstrating an in-plane targeting error of 1.5[Formula: see text]mm. The robotic system has been approved by the institutional review board (IRB) for clinical trials. A preliminary clinical study is conducted with the patient consent, demonstrating the targeting errors at two biopsy target sites to be 4.0[Formula: see text]mm and 3.7[Formula: see text]mm, which is sufficient to target a clinically significant tumor foci. First-in-human trials to evaluate the system’s effectiveness and accuracy for MR image-guided prostate biopsy are underway.
40
3D Slicer as an image computing platform for the Quantitative Imaging Network
Andriy Fedorov, Reinhard Beichel, Jayashree Kalpathy–Cramer et al. · Magnetic Resonance Imaging · 2012 · 8.9K citations · Full text
Arts, Computer Science, Radiology +12
Real-time MRI-TRUS fusion for guidance of targeted prostate biopsies
Sheng Xu, Jochen Kruecker, Barış Türkbey et al. · Computer Aided Surgery · 2008 · 299 citations
MR Imaging–guided Prostate Biopsy with a Closed MR Unit at 1.5 T: Initial Results