Journal of Computer and Communications · 2016 · 17 citations · 14 references
EngineeringElectrical EngineeringInstrumentationMicroelectronicsMicrofabricationAntenna MeasurementCalibrationDevice CharacterizationEquipment CalibrationMicrowave MeasurementsMicrowave Network AnalysisCalibration TechniquesDevice CalibrationInstrumentation CalibrationMisalignment EffectsOn-wafer MeasurementOn-wafer Measurement TechniquesProbe CalibrationTest Equipment Calibration
A fully automatic setup for on-wafer contact probing will be presented. This setup consists of six automatable nano positioning axes used as tool holder and a sample holder. With this setup a fully automatic one-port SOL calibration for a Vector Network Analyzer is done. Furthermore a fully automated on-wafer contact probing is performed. Afterwards, the effects of a misalignment of the three tips of a GSG-probe are examined. Additionally the error on the calibration is calculated to determine its effect on the measurement. The results show, that a misalignment of the probe has a high impact on the measurement of the VNA. Hence a fully automated on-wafer probing presented in this paper is a good way to detect these misalignments and correct them if necessary.
14
Gary Bradski · Medical Entomology and Zoology · 2000 · 3.1K citations
E.P. Vandamme, Dominique Schreurs, G. Van Dinther · IEEE Transactions on Electron Devices · 2001 · 256 citations
Comparison of On-Wafer Calibrations
Dylan F. Williams, Roger B. Marks, Andrew Davidson · 1991 · 224 citations