DOAJ (DOAJ: Directory of Open Access Journals) · 2008 · 16 citations · 18 references
Open access
Natural SciencesChemistryFood ScienceAnalytical ChemistryFood AnalysisSpectroscopic AnalysisAbsorption SpectroscopyBioanalytical Method ValidationInfrared SpectroscopyMetabolite ExtractionOptical SpectroscopySpectroscopic Food AnalysisAbsorption SpectrumFourier Transform AnalysisFourier Transform Infrared SpectroscopyInfrared Absorption SpectraInfrared Absorption SpectroscopyInfrared SpectraInfrared Spectroscopic AnalysisPrecision
"In this work a flow analysis-Fourier transform infrared spectrometric (FA-FTIR) method for the determination of caffeine in coffee was developed. The method involves the extraction, in batch, of caffeine from solid samples of roasted coffee with hot water (80°C), on-line retro-extraction of caffeine in chloroform and direct analysis of the organic phase by FTIR spectrometry. The measurement criterion selected for the quantification of caffeine was the absolute peak-valley value (1668-1649 cm–1) of the first order derivate of the absorption spectrum, corresponding to the 1659 cm–1 band. The method provides a limit of detection (3) of 0.001% (w/v) caffeine, a relative standard deviation corresponding to the whole procedure, until the sample preparation, of 1.2% ([Caffeine] 0.05% (w/v), n = 5) and a sample throughput (triplicate) of approximately 12 samples h–1. The method was validated by means of standard addition, recovery and inter-day precision studies and by comparative analysis of real samples with the official chromatographic-spectrophotometric procedure. The method was applied to the analysis of commercial samples, obtaining a caffeine content ranging from 0.60 to 1.16% (w/w)."
18