Use of an Autosampler for Dynamic Headspace Extraction of Volatile Compounds from Grains and Effect of Added Water on the Extraction

Mohan Ram, Larry M. Seitz, R. Rengarajan

Journal of Agricultural and Food Chemistry · 1999 · 15 citations · 9 references

Concepts

Abstract

An autosampler attached to a purge and trap instrument was used to aid routine analyses of grain samples for volatile compounds associated with off-odors. Trapped volatiles were transferred to a gas chromatograph/mass spectrometer instrument for separation and detection. Dynamic extraction of volatiles from approximately 18 g of whole grain at 80 degrees C was accomplished by purging helium through a sample vial with a Teflon-lined septum on each end. The autosampler automatically added internal standard to the sample before purging began, which required the addition of 1 mL of water for complete transfer of the standard to the sample. The added water enhanced extraction of 1-octen-3-ol, 1-octen-3-one, and some other compounds from soybeans but not from starchy grains such as corn and wheat. Addition of a free radical scavenger, such as citric acid, greatly diminished the recovery of 1-octen-3-ol and 1-octen-3-one from soybeans.