Measurement Science and Technology · 1996 · 282 citations · 9 references
EngineeringMechanical EngineeringMicroelectromechanical SystemsMicrofabricationSurface ScienceAtomic Force MicroscopyMicromachiningNanomechanical PropertiesNanomechanicsScanning Probe MicroscopyAtomic Force Microscopy CalibrationAtomic Force Microscopy MetrologyCantilever MechanicsForce-distance CurvesHeterodyne DetectionHeterodyne InterferometrySpring CharacteristicsSpring Constant
Cantilevers fabricated by means of micromachining techniques are usually used for atomic force microscopy. In this paper, the spring constant of an atomic force microscope (AFM) cantilever is determined by using a large-scale cantilever. Since the spring constant of the large-scale cantilever is calibrated accurately, the spring constant of the AFM cantilever is determined precisely by measuring the deflections of both cantilevers simultaneously using heterodyne interferometry. The slope of the force curve gives the spring constant of the AFM cantilever. It is not necessary to measure the dimensions of the AFM cantilever in the proposed method. Although this method is simple, the spring constant of the AFM cantilever is obtained accurately.
9
Interpretation of force curves in force microscopy
N. A. Burnham, Richard J. Colton, H. M. Pollock · Nanotechnology · 1993 · 304 citations