2020 · 10 citations · 34 references
MedicineRehabilitation MedicineRobotics EngineeringBiomechanicsProstheticsRoboticsImpedance ControlLower-limb ProstheticsProsthetic Control SystemsProsthetic RoboticsRehabilitation RoboticsTransfemoral ProstheticsAmputee GaitBio-inspired ActuationLower-limb Prosthesis ControlLower-limb Prosthetic ControlPowered ProsthesesPowered Prosthetic JointsProsthetic ControlProsthetic Limb Control
This paper reports the development and preliminary validation of a bio-inspired controller using an adaptive oscillator and artificial primitives for a powered ankleknee prosthesis, namely the CYBERLEGsPlusPlus Gammaprosthesis. The proposed controller combines a feed-forward reference torque component with a feedback impedance-based torque component for the prosthetic knee and ankle joints. A treadmill walking experiment was conducted with a transfemoral amputee to assess the relevance of the feed-forward torque component, compared to a pure feedback control approach. Results showed that adding a feed-forward torque component to both prosthetic joints allowed a decrease of the feedback gains, and thus a more compliant behavior of the prosthesis. This combined condition was well-perceived by the pilot transfemoral amputee.
34
Human-in-the-loop optimization of exoskeleton assistance during walking
Juanjuan Zhang, Pieter Fiers, Kirby Ann Witte et al. · Science · 2017 · 1K citations
Energy cost of walking of amputees: the influence of level of amputation.
Waters Rl, Jacquelin Perry, Daniel Antonelli et al. · PubMed · 1976 · 879 citations
Energy cost of walking of amputees
Waters Rl, Jacquelin Perry, Daniel Antonelli et al. · Journal of Bone and Joint Surgery · 1976 · 833 citations