Determination of bioactive nonvolatile ginger constituents in dietary supplements by a rapid and economic HPLC method: Analytical method development and single-laboratory validation

Hong You, Bailey Ireland, Michael Moeszinger, Haoshu Zhang, Laura Snow, Scott Krepich, Vivian Takagawa

Talanta · 2018 · 45 citations · 14 references

DOIFull text

Open access

Concepts

Abstract

< 8%). Among 10 commercial ginger-containing samples that we screened using this method, the results were 80-123% of the products' labeling value. The HPLC running time was successfully shortened from 29 min (USP method) and 40 min (ISO method) to 12 min without the need of using an expensive Mass Spectrometer for analyte separation. The method is the first method that meets all AOAC SMPR 2017.12 requirements and therefore has the potential to be adopted as a consensus industrial reference method for meeting FDA's cGMP Compliance for the manufacture and quality control of dietary supplements and ingredients.