2005 · 36 citations · 2 references
EngineeringBiomedical EngineeringMechanical EngineeringActuatorsMicroelectromechanical SystemsMicrofabricationImplantable Biomedical DevicesMicromachiningShape Memory Alloy ActuatorsBendingBiomedical MicroactuatorsDirect HeatingIndirect HeatingMicroelectromechanical ActuatorsMiniature ActuatorsShape-memory Alloy ActuationSilicon Micromachining
Wc dcvclopcd a new prototype active catheter with multi-link structure based on silicon micromachining. Many links are connected with joints made of Shape Memory Alloy (SMA) actuators. The active catheter can be used for medical purposes as an interventional therapy. As this has multi-link structure, it can bend actively on demand and conform to the shape of narrow blood vessels. The outer diameter of the fabricated active cathctcr is 2.8. Micro coil of SMA is used as the actuator for the joint. Two heating methods, direct heating and indirect heating of SMA was examined. Indirect heating by surrounding nickel thin film heater is compatible with built-in circuit for driving the SMA actuator. We used silicon batch process for the fabrication of the link to minimize assembly work. UV (Ultra Violet) curing resin was used to fix the actuator and to make electrical contact. Bending motions, bending angle, force, and frequency response of the active catheter were measured.
2