Styrene and Ethylbenzene Migration from Polystyrene into Dairy Products by Dynamic Purge‐and‐Trap Gas Chromatography

J. Ehret‐Henry, Violette Ducruet, Anna Maria Luciani, Alexandre Feigenbaum

Journal of Food Science · 1994 · 31 citations · 9 references

Concepts

Abstract

ABSTRACT Ethylbenzene and styrene were determined by an automated dynamic headspace system with a purge‐and‐trap injector. Hydrocarbons were purged with He into a cold trap and separated by GC. The procedure. was applied to dairy products without sample preparation. Preliminary He treatment, temperature of condenser, and of the analytical cell were important. Results of styrene migration from the package into dairy products, including yogurts, cream and chocolate desserts, were well below 5μg/kg.

References

9