System design of an Insertable Robotic Effector Platform for Single Port Access (SPA) Surgery
149
Citations
23
References
2009
Year
Unknown Venue
Robotic SystemsEngineeringTeleoperationMinimally Invasive ProcedureSurgical ScienceSurgeryBiomedical EngineeringSystem DesignKinematicsSurgical PlanningRoboticsComputer-assisted SurgeryMechanical DesignRobotic TechnologyMedicineMechatronicsMedical RobotSingle Port AccessRobotic SurgeryMechanical SystemsRobot-assisted SurgeryIrep RobotMedical DevicesPreliminary Kinematic AnalysisSurgical Innovation
This paper presents a novel design and preliminary kinematic analysis of an insertable robotic effector platform (IREP) for single port access (SPA) surgery. The IREP robot can be deployed into body cavity through a Ø15 mm skin incision to perform SPA procedures. It consists of two snake-like continuum robots as slave surgical assistants for tissue manipulation, two parallelogram mechanisms for the continuum robots' placement, and one controllable stereo vision module with integrated light source for depth perception and tool tracking. Design considerations and alternatives, calculations and preliminary simulations of this 17-DoF surgical robotic system are presented in this paper. The overall control system hierarchy for tele-manipulation using the IREP robot is also presented.