Review of Scientific Instruments · 1999 · 100 citations · 13 references
EngineeringMechanical EngineeringInstrumentationMicroelectromechanical SystemsVibrationAtomic Force MicroscopyPiezoelectric ActuationPiezoelectric ActuatorsPiezoelectric MaterialsScanning Probe MicroscopyDynamic Force MicroscopyLow-temperature Scanning Probe MicroscopyMagnetic Force MicroscopyMicroscope SetupOscillator DevicesQuartzScanned Probe Applications
A scanning force microscope was implemented operating at temperatures below 4.2 K and in magnetic fields up to 8 T. Piezoelectric quartz tuning forks were employed for nonoptical tip–sample distance control in the dynamic operation mode. Fast response was achieved by using a phase-locked loop for driving the mechanical oscillator. Possible applications of this setup for various scanning probe techniques are discussed.
13