Journal of Vacuum Science & Technology B Microelectronics and Nanometer Structures Processing Measurement and Phenomena · 1994 · 29 citations · 0 references
Natural SciencesMaterials EngineeringMaterials ScienceNanomaterialsNanotechnologySurface ScienceElectron MicroscopyScanning Probe MicroscopySurface PhysicsTransmission Electron MicroscopyAnodic BondingField EmissionHigh-resolution Electron MicroscopyMicroscope AlignmentScanning Transmission Electron MicroscopySiliconTip AlignmentTip Characteristics
Recent advances in microcolumn design and fabrication have led to a significant improvement in the resolution obtained with a new 3.5 mm long microcolumn. At an electron energy of 1 keV and 2 mm working distance, the beam diameter, measured by the signal rise-time while scanning over a sharp edge, is 40 nm and the resolution observed in scanning transmission electron microscope images is about 30 nm. These experimental results agree well with results of the electron-optical modeling. All lenses and apertures in the column are made from silicon membranes and assembled using a novel ultrahigh vacuum compatible, multilayer anodic bonding technique. The scanning tunneling microscope aligned field-emission source uses an oxygen processed, 50 nm radius, cold 〈111〉 W field-emission tip. Good emission stability for scanning electron microscope operation has been achieved with a few percent of root-mean-square current fluctuation over about 30 min. With periodic tip flashes, the system has been operated reliably over several weeks. A novel in situ tip alignment technique with respect to the extractor electrode has been developed. This technique also provides information about the tip emission characteristics during column operation.