Publication | Open Access
An enhanced marker pattern that achieves improved accuracy in surgical tool tracking
29
Citations
19
References
2021
Year
Cylindrical MarkersEngineering3D Pose EstimationSurgical ToolField RoboticsEnhanced Marker PatternSurgeryBiomedical EngineeringMedical InstrumentationImage AnalysisObject TrackingComputational ImagingEx Vivo DataKinematicsSurgical PlanningRadiologyMachine VisionComputer-assisted SurgeryRobotic TechnologyMedicineSurgical TrainingImage GuidanceMedical Image ComputingImproved AccuracyComputer VisionSurgical Motion AnalysisVideo AnalysisEye TrackingExtended RealitySurgical Tool TrackingRoboticsSurgical Innovation
In computer assisted interventions (CAI), surgical tool tracking is crucial for applications such as surgical navigation, surgical skill assessment, visual servoing, and augmented reality. Tracking of cylindrical surgical tools can be achieved by printing and attaching a marker to their shaft. However, the tracking error of existing cylindrical markers is still in the millimetre range, which is too large for applications such as neurosurgery requiring sub-millimetre accuracy. To achieve tool tracking with sub-millimetre accuracy, we designed an enhanced marker pattern, which is captured on images from a monocular laparoscopic camera. The images are used as input for a tracking method which is described in this paper. Our tracking method was compared to the state-of-the-art, on simulation and ex vivo experiments. This comparison shows that our method outperforms the current state-of-the-art. Our marker achieves a mean absolute error of 0.28 [mm] and 0.45 [°] on ex vivo data, and 0.47 [mm] and 1.46 [°] on simulation. Our tracking method is real-time and runs at 55 frames per second for 720×576 image resolution.
| Year | Citations | |
|---|---|---|
Page 1
Page 1