Rapid determination of tannin in Danshen and Guanxinning injections using UV spectrophotometry for quality control

Hongxia Huang, Yuanyuan Lv, Xiaoyi Sun, Shuangshuang Fu, Xue-Fang Lou, Zengmei Liu

Journal of Innovative Optical Health Sciences · 2018 · 11 citations · 11 references

DOIFull text

Open access

Concepts

Abstract

A technique for the determination of tannin content in traditional Chinese medicine injections (TCMI) was developed based on ultraviolet (UV) spectroscopy. Chemometrics were used to construct a mathematical model of absorption spectrum and tannin reference content of Danshen and Guanxinning injections, and the model was verified and applied. The results showed that the established UV-based spectral partial least squares regression (PLS) tannin content model performed well with a correlation coefficient ([Formula: see text]) of 0.952, root mean square error of calibration (RMSEC) of 0.476[Formula: see text][Formula: see text]g/ml, root mean square error of validation (RMSEV) of 1.171[Formula: see text][Formula: see text]g/ml, and root mean square error of prediction (RMSEP) of 0.465[Formula: see text][Formula: see text]g/ml. Pattern recognition models using linear discriminant analysis (LDA) and [Formula: see text] nearest neighbor ([Formula: see text]-NN) classifiers based on UV spectrum could successfully classify different types of injections and different manufacturers. The established method to measure tannin content based on UV spectroscopy is simple, rapid and reliable and provides technical support for quality control of tannin in Chinese medicine injections.

References

11