Surface and Interface Analysis · 2018 · 26 citations · 10 references
Natural SciencesMaterials EngineeringMaterials ScienceSemiconductor DevicesSpectroscopic AnalysisSpectroscopyDevice CharacterizationPhotoelectron SpectroscopySemiconductor SurfacesSurface CharacterizationX-ray Photoelectron SpectroscopyX-ray SpectroscopyChemical Shift AnalysisHard X-ray SpectroscopyHigh-resolution X-ray SpectroscopyLaboratory X-ray SpectroscopyPhoton EnergyQuantitative Surface AnalysisSpectrometer DesignX-ray Chemical Shift Analysis
Applications of laboratory hard X‐ray photoelectron spectroscopy on buried interfaces in devices are presented. We use a novel spectrometer fitted with a monochromated CrKα source (photon energy: 5414.9 eV) and a high‐voltage analyzer. Elements buried at depths as deep as 25 nm underneath various overlayers such as Al/Ta and Pt/Ti are detected and quantified from survey spectra, with chemical shifts accessible from high‐resolution scans. Different analysis conditions towards optimizing the information depth are presented and discussed.
10
Hiroshi Shinotsuka, Shigeo Tanuma, C. J. Powell et al. · Surface and Interface Analysis · 2015 · 369 citations
Natural Sciences, Chemistry, Physics +13