Achieving high effective<i>Q</i>-factors in ultra-high vacuum dynamic force microscopy

Jannis Lübbe, L. Tröger, Stefan Torbrügge, Ralf Bechstein, Christoph Richter, Angelika Kühnle, Michael Reichling

Measurement Science and Technology · 2010 · 51 citations · 30 references

Concepts

Abstract

The effective Q-factor of the cantilever is one of the most important figures-of-merit for a non-contact atomic force microscope (NC-AFM) operated in ultra-high vacuum (UHV). We provide a comprehensive discussion of all effects influencing the Q-factor and compare measured Q-factors to results from simulations based on the dimensions of the cantilevers. We introduce a methodology to investigate in detail how the effective Q-factor depends on the fixation technique of the cantilever. Fixation loss is identified as a most important contribution in addition to the hitherto discussed effects and we describe a strategy for avoiding fixation loss and obtaining high effective Q-factors in the force microscope. We demonstrate for room temperature operation, that an optimum fixation yields an effective Q-factor for the NC-AFM measurement in UHV that is equal to the intrinsic value of the cantilever.

References

30