2014 · 31 citations · 14 references
EngineeringBiomedical EngineeringRobotics EngineeringMedical RoboticsRobot KinematicsRoboticsCatheter AblationDifferential KinematicsMagnetic ActuatorsMedical Robot DynamicsMedical Robot KinematicsSurgical RoboticsCatheter-based RoboticsDeflection CalculationKinematic Modeling Of Minimally Invasive RobotsProof Of Concept PrototypesRobot Kinematic ModelingRobotic CathetersThree-dimensional Kinematic AnalysisTorsion
This paper presents the three dimensional kinematic modeling of a novel steerable robotic ablation catheter system. The catheter, embedded with a set of current-carrying micro-coils, is actuated by the magnetic forces generated by the magnetic field of the MRI scanner. This paper develops a 3D model of the MRI actuated steerable catheter system by using finite differences approach. For each finite segment, a quasi-static torque-deflection equilibrium equation is calculated using beam theory. By using the deflection displacements and torsion angles, the kinematic modeling of the catheter system is derived. The proposed models are evaluated by comparing the simulation results of the proposed model with the experimental results of a proof-of-concept prototype.
14
Advanced mechanics of materials
A.R. Flint · Journal of the Mechanics and Physics of Solids · 1953 · 240 citations
New approaches to catheter navigation for interventional radiology simulation
Christian Duriez, S. Cotin, Julien Lenoir et al. · Computer Aided Surgery · 2006 · 96 citations · Full text