IEEE Electron Device Letters · 2005 · 26 citations · 10 references
EngineeringElectrical EngineeringPhysicsOpticsOptoelectronicsSemiconductor DevicesCalibrationDevice CharacterizationEquipment CalibrationLaser DiodesTemperature MeasurementThermal Properties Of SemiconductorsCalibration Of Optical InstrumentsCatastrophic Optical DamageOptical ThermometryPeltier EffectThermal Calibration
In this letter, we present a thermoreflectance setup specially designed for the study of the temperature variations of the output facet of a laser diode. Indeed, the temperature of the laser diode is controlled by a Peltier element and the device under test is used as a temperature sensor. We propose a calibration procedure based on electrical measurements combined with optical ones; it leads to the determination of thermoreflectance coefficients and then to absolute temperature variations on a running laser diode. We can hence ensure a proper running of the diode and avoid its catastrophic optical facet damage.
10
Stéphane Grauby, Benoı̂t C. Forget, Stéphane Holé et al. · Review of Scientific Instruments · 1999 · 137 citations
Peter-Wolfgang Epperlein Peter-Wolfgang Epperlein · Japanese Journal of Applied Physics · 1993 · 98 citations