Current Analytical Chemistry · 2009 · 13 citations · 15 references
AgricultureChemical EngineeringFood ScienceAnalytical ChemistryFood AnalysisTransport PhenomenaAnalytical ChromatographyJuice ProcessingSugar AnalysisSugar ProcessingAnalytical PrecisionFood Sugar AnalysisHydrolysisLaminar Flow RegimesPumping SystemsRadial TransportReducing Sugar DeterminationSucrose Quantification
Laminar and pulsed flows typical of multi-commuted and multi-pumping flow systems, were evaluated in relation to analytical procedures carried out at high temperatures. As application, the spectrophotometric determination of total reducing sugars (TRS, hydrolyzed sucrose plus reducing sugars) in sugar-cane juice and molasses was selected. The method involves in-line hydrolysis of sucrose and alkaline degradation of the reducing sugars at about 98 °C. Better results were obtained with pulsed flows, due to the efficient radial mass transport inherent to the multi-pumping flow system. The proposed system presents favorable characteristics of ruggedness, analytical precision (r.s.d. < 0.013 for typical samples), stability (no measurable baseline drift during 4-h working periods), linearity of the analytical curve (r > 0.992, n = 5, 0.05 - 0.50% w/v TRS) and sampling rate (65 h-1). Results are in agreement with ion chromatography. Keywords: Multi-pumping flow analysis, Pulsed flow, Laminar flow, Heat transference, Total reducing sugars, Sugar-cane juice and molasses, Spectrophotometry
15
Boaventura F. Reis, María Fernanda Giné, Elias A.G. Zagatto et al. · Analytica Chimica Acta · 1994 · 320 citations