Journal of Micromechanics and Microengineering · 2001 · 32 citations · 6 references
EngineeringMaterials EngineeringMaterials ScienceHeat TransferMicrofabricationSensorsHeat ConductionTemperature MeasurementTemperature SensorsThermal SensorsEnergy TransferHeat Flux MeasurementHeat Flux SensingTemperature SensingThermal SensingThermal Sensor CalibrationThermal TransferThermocouples
A micro heat flux sensor which can measure the thermal energy transfer per unit area has been designed, fabricated, and calibrated in a convective environment. The sensor which is based on a circular foil gage is composed of thermal paths and a thermopile. The thermal path is made in a LIGA-like process of SU-8 high aspect ratio microstructures and electroplated copper layers. The thermopile, a series of thermocouples, is used to amplify the output signal as a thermometer. When the sensor is placed on a high-temperature wall, heat flux from the wall flows through thermal paths and drains out to the environment, producing a temperature difference along its paths. Heat flux is obtained by calibrating this temperature difference in the thermopile of Ni-Cr or Al-Chromel pairs. The sensitivity of the heat flux sensor of Ni-Cr and Al-Chromel pairs is in the range of 0.1-2.0 and 0.4-2.0 muV mW(-1) cm(-2), respectively, in the heat flux range of 0-180 mW cm(-2).
6
High-frequency response heat-flux gauge
A. H. Epstein, G. R. Guenette, R. J. G. Norton et al. · Review of Scientific Instruments · 1986 · 106 citations
Engineering, Materials Engineering, Mechanical Engineering +17
A Transducer for the Measurement of Heat-Flow Rate
Robert Gardon · Journal of Heat Transfer · 1960 · 67 citations