Secondary ion mass spectrometry depth profiling of ultralow-energy ion implants: Problems and solutions

J. G. M. van Berkum, E. J. H. Collart, K. Weemers, D. J. Gravesteijn, K. Iltgen, A. Benninghoven, E. Niehuis

Journal of Vacuum Science & Technology B Microelectronics and Nanometer Structures Processing Measurement and Phenomena · 1998 · 16 citations · 2 references

DOIFull text

Open access

Concepts

Abstract

The use of oxygen flooding and an amorphous Si capping for secondary ion mass spectrometry (SIMS) depth profiling of ultralow-energy implants is discussed in terms of the accuracy of the depth and concentration scale. Further, the advantages of a lower primary ion energy and the simultaneous mass registration on a time-of-flight-SIMS with dual beam mode is illustrated. Finally, the effects of ultrahigh top concentrations on the tail of implantation profiles is investigated.