The international archives of the photogrammetry, remote sensing and spatial information sciences/International archives of the photogrammetry, remote sensing and spatial information sciences · 2016 · 67 citations · 13 references
EngineeringMechatronics EngineeringPhotonicsSensorsThermal AnalysisCalibrationRange EstimationSensor CalibrationSpatial CalibrationAccuracyGeometric CalibrationLasersLidar InstrumentationLidar SensorsLidar SystemsLong-term StabilityTemporal Stability
Abstract. We report on a calibration and stability analysis of the Velodyne VLP-16 LiDAR scanner. The sensor is evaluated for long-term stability, geometric calibration and the effect of temperature variations. To generalize the results, three separate VLP-16 sensors were examined. The results and conclusions from the analysis of each of the individual sensors was similar. We found that the VLP-16 showed a consistent level of performance, in terms of range bias and noise level over the tested temperature range from 0–40 °C. A geometric calibration was able to marginally improve the accuracy of the VLP-16 point cloud (by approximately 20%) for a single collection, however the temporal stability of the geometric calibration negated this accuracy improvement. Overall, it was found that there is some long-term walk in the ranging observations from individual lasers within the VLP-16, which likely causes the instability in the determination of geometric calibration parameters. However, despite this range walk, the point cloud delivered from the VLP-16 sensors tested showed an accuracy level within the manufacturer specifications of 3 cm RMSE, with an overall estimated RMSE of range residuals between 22 mm and 27 mm.
13
Calibration between Color Camera and 3D LIDAR Instruments with a Polygonal Planar Board
Yoonsu Park, Seokmin Yun, Chee Sun Won et al. · Sensors · 2014 · 232 citations · Full text
Faraz M. Mirzaei, Dimitrios G. Kottas, Stergios I. Roumeliotis · The International Journal of Robotics Research · 2012 · 161 citations