Rapid Communications in Mass Spectrometry · 2003 · 15 citations · 10 references
PharmacyPharmacologyToxicologyAnalytical ToxicologyMass Spectrometric AnalysisPharmaceutical ToxicologyBioanalytical Method ValidationDrug DispositionLiquid ChromatographyMass SpectrometryMetabolite ExtractionMetabolite QuantificationHigh-performance Liquid Chromatography-mass SpectrometryLiquid Chromatography-mass SpectrometryLiquid Chromatography-mass Spectrometry AnalysisLower Limit Of QuantificationLower Limit Of QuantitationMass Spectrometric Drug AnalysisPrecisionSweetener Analysis
The main toxicological concern of stevioside, a highly potent sweetener from S. rebaudiana, is its main metabolite, steviol. To determine very low levels of steviol in in vivo experiments, a sensitive liquid chromatography/atmospheric pressure chemical ionization mass spectrometry (LC/APCI-MS) method was developed for quantifying steviol in rat plasma after oral administration of a single dose of stevioside (0.5 g/kg). The sample preparation uses liquid-liquid extraction with tert-butyl methyl ether in an acidic environment. The retention time of steviol was 10.5 min. The assay was linear over the range 2-1000 ng/mL with a lower limit of detection of 1 ng/mL. The intra- and inter-day precision were <5 and <7%, respectively, and the accuracy was in the range 95-108%. The steviol concentration profile in rat plasma was determined.
10
Bioanalytical Method Validation—A Revisit with a Decade of Progress
Vinod P. Shah, Kamal K. Midha, John W. A. Findlay et al. · Pharmaceutical Research · 2000 · 1.6K citations
G. G. Birch · Food Chemistry · 1992 · 145 citations