Sensors · 2014 · 32 citations · 17 references
EngineeringBioengineeringElectrical EngineeringElectrochemical EngineeringElectrochemistryMicroelectronicsMicroelectrode ArraysMicrofluidic DevicesMicrofluidic SystemsMicrofluidicsReference ElectrodesWorking ElectrodesArray ElectrodesElectrochemical CharacterizationElectrode FabricationIntegrating Microfluidics With Sensing ElementsLift-offMicrofabrication Of Electrode ArraysReference Electrode DesignSilicon
In this paper we demonstrate the fabrication and electrochemical characterization of a microchip with 12 identical but individually addressable electrochemical measuring sites, each consisting of a set of interdigitated electrodes acting as a working electrode as well as two circular electrodes functioning as a counter and reference electrode in close proximity. The electrodes are made of gold on a silicon oxide substrate and are passivated by a silicon nitride membrane. A method for avoiding the creation of high edges at the electrodes (known as lift-off ears) is presented. The microchip design is highly symmetric to accommodate easy electronic integration and provides space for microfluidic inlets and outlets for integrated custom-made microfluidic systems on top.
17
Gold cleaning methods for electrochemical detection applications
Lee MacKenzie Fischer, Maria Tenje, Arto Heiskanen et al. · Microelectronic Engineering · 2008 · 330 citations