Analytical Chemistry · 1994 · 147 citations · 17 references
Natural SciencesBioengineeringMaterials ScienceBiocompatibilityElectroanalytical MethodsElectrochemistryCoatingsImplant MaterialsImplantable BiomaterialsImplantable Biomedical DevicesSurface Modification Of ImplantsDegradationImplant CoatingsImplant DegradationResearch ArticlesSurface Coating BiocompatibilitySurface Coatings For Implants
To assess the effect of the biological response to implanted Ag/AgCl reference electrodes on the electrode stability, uncoated and polymer-coated Ag/AgCl electrodes were implanted subcutaneously in rats. After 1 week of implantation, uncoated Ag/AgCl electrode potentials, measured in 0.1 M KCl, shifted by about -180 mV, and both voltammetry and electron microscopy showed that all the AgCl was removed. The electrodes could be significantly protected by coating with polyurethane or a perfluorinated ionomer (Nafion) cured at 120 degrees C for 1 h. Electron micrographs showed the 120 degrees C cured Nafion and polyurethane coatings remained intact over 2 weeks of implantation. Following 2 weeks of implantation the cured, Nafion-coated electrodes' potentials were shifted by -15 +/- 7 mV relative to the initial values. Voltammetry showed that they were still not polarizable. The current densities obtained with the coated reference electrodes are sufficient for their use as counter/pseudoreference electrodes with implantable two-electrode glucose sensor systems. The tissue response to coated electrodes was minimal in comparison to the response to uncoated reference electrodes.