Biomedical Chromatography · 2009 · 15 citations · 16 references
PharmacyPharmacokineticsPharmacologyChromatographic AnalysisFood AnalysisMass Spectrometric AnalysisBioanalytical Method ValidationLiquid ChromatographyMass SpectrometryConcentrationElectrospray IonizationInternal Standard AnalysisLiquid Chromatography-mass SpectrometryLiquid Chromatography-mass Spectrometry AnalysisLiquid Chromatography-mass Spectrometry Method Development
A sensitive and specific liquid chromatographic-electrospray ionization mass spectrometric method was developed for quantification of salvianolic acid B in rat plasma with resveratrol as the internal standard. The analytes were separated on a reversed-phase column with acetonitrile (40%) and water (60%) containing 0.75% formic acid as mobile phase at a flow rate of 1 mL/min. Liquid-liquid extraction was adopted for the sample preparation, and the analytes were determined using electrospray negative ionization mass spectrometry in the selective monitoring mode. The method was validated over the concentration range 0.1-40 microg/mL using 0.1 mL of plasma with coefficients of correlation >0.999. The intra- and inter-day precisions of analysis were <10%, and accuracy ranged from 94 to 101%. This method was successfully applied to a pharmacokinetics of salvianolic acid B in rats.
16