Longitudinal bulk acoustic mass sensor

Jan Harry Hales, J. Teva, Anja Boisen, Zachary J. Davis

Applied Physics Letters · 2009 · 15 citations · 15 references

DOIFull text

Open access

Concepts

Abstract

A polycrystalline silicon longitudinal bulk acoustic cantilever is fabricated and operated in air at 51 MHz. A mass sensitivity of 100 Hz/fg (1 fg=10−15 g) is obtained from the preliminary experiments where a minute mass is deposited on the device by means of focused ion beam. The total noise in the currently applied measurement system allows for a minimum detectable mass of 0.5 fg in air.