Microstructured Apertures in Planar Glass Substrates for Ion Channel Research

Niels Fertig, Michael George, Michèle Klau, Christine Meyer, A. Tilke, Constanze Sobotta, Robert H. Blick, Jan C. Behrends

Receptors and Channels · 2003 · 52 citations · 4 references

Concepts

Abstract

We have developed planar glass chip devices for patch clamp recording. Glass has several key advantages as a substrate for planar patch clamp devices. It is a good dielectric, is well-known to interact strongly with cell membranes and is also a relatively in-expensive material. In addition, it is optically neutral. However, microstructuring processes for glass are less well established than those for silicon-based substrates. We have used ion-track etching techniques to produce micron-sized apertures into borosilicate and quartz-glass coverslips. These apertures, which can be easily produced in arrays, have been used for high resolution recording of single ion channels as well as for whole-cell current recordings from mammalian cell lines. An additional attractive application that is greatly facilitated by the combination of planar geometry with the optical neutrality of the substrate is single-molecule fluorescence recording with simultaneous single-channel measurements.

References

4