Design, Fabrication, and Assembly of an Optical Biosensor Probe Package for OCT (Optical Coherence Tomography) Application

C.S. Premachandran, Ahmad Khairyanto, Kelvin Chen Wei Sheng, J. Singh, Jason Teo, Yingshun Xu, C. Nanguang, Colin J. R. Sheppard, Malini Olivo

IEEE Transactions on Advanced Packaging · 2009 · 11 citations · 6 references

Concepts

Abstract

A miniaturized optical bioprobe package is developed using a 3-D micromirror and is tested for bio-imaging application. A silicon optical bench is designed and micromachined to assemble the fiber, lens, and the 3-D micromirror device. A 45 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">deg</sup> angle trench is used to place the micromirror to achieve larger scanning range. Trace lines are formed on the optical bench and are connected to silicon micromirror using solder. A GRIN lens with lower numerical aperture has been used to focus the optical beam onto the micromirror. The bioprobe is packaged and is tested in a time domain optical coherence tomography (OCT) setup and optical image is obtained for plant tissue.

References

6