EngineeringMicroscopyMechanical EngineeringMicroelectromechanical SystemsBiomedical EngineeringMicro-optical ComponentEndoscopic ImagingRotational TorqueMems ScannerMicro-electromechanical SystemTissue ImagingInstrumentationRotational EnergyRadiologyMedical ImagingOphthalmologyMechatronicsBiophotonicsOptical ImagingMicrofabricationBiomedical ImagingOptical Coherence TomographyMedicineTomography
A novel endoscopic optical coherence tomography probe was designed and constructed with a 1.9-mm microelectromechanical system (MEMS) motor. The new MEMS endoscopic probe design eliminates the need to couple the rotational energy from the proximal to the distal end of the probe. Furthermore, the endoscopic probe's sheath and fiber have the advantages of having a much smaller diameter and being more flexible than traditional endoscopes since no reinforcement is needed to couple the rotational torque. At the distal end, a prism mounted on a micromotor deflects the light rays to create a transverse circular-scanning pathway. Because our MEMS scanner does not require the coupling of a rotational single-mode fiber, a high scanning speed is possible while eliminating unstable optical signals caused by nonuniform coupling.
7
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.5K citations
Scanning single-mode fiber optic catheter–endoscope for optical coherence tomography
G. J. Tearney, Mark E. Brezinski, James G. Fujimoto et al. · Optics Letters · 1996 · 354 citations
Electrostatic micromachine scanning mirror for optical coherence tomography
Jason M. Zara, Siavash Yazdanfar, K. Divakara Rao et al. · Optics Letters · 2003 · 105 citations