Endoscopic optical coherence tomography based on a microelectromechanical mirror
267
Citations
12
References
2001
Year
Microelectromechanical MirrorLateral LightTissue ImagingEngineeringOphthalmologyMedical ImagingMicroscopyMedicineBladder TissueBiomedical ImagingBiomedical OpticOptical Coherence TomographyBiophotonicsBiomedical EngineeringOct ScopeEndoscopic ImagingOptical ImagingTomography
An endoscopic optical coherence tomography (OCT) system based on a microelectromechanical mirror to facilitate lateral light scanning is described. The front-view OCT scope, adapted to the instrument channel of a commercial endoscopic sheath, allows real-time cross-sectional imaging of living biological tissue via direct endoscopic visual guidance. The transverse and axial resolutions of the OCT scope are roughly 20 and 10.2mum, respectively. Cross-sectional images of 500x1000 pixels covering an area of 2.9 mmx2.8 mm can be acquired at ~5 frames/s and with nearly 100-dB dynamic range. Applications in thickness measurement and bladder tissue imaging are demonstrated.