Review of Scientific Instruments · 2006 · 13 citations · 26 references
Natural SciencesBiophysicsPhysicsMolecular BiophysicsSingle-molecule BiophysicsStructural BiophysicsAtomic Force MicroscopyBiomolecular ThermodynamicsScanning Probe MicroscopySingle-molecule ImagingSingle-molecule MeasurementsTemperature MeasurementForce-dependent Protein UnfoldingHeat Sink DesignMechanical Unfolding PathwaysTemperature-dependent Structural ChangesThermoelectric CoolingThermoelectric Module DesignThermoelectric PropertiesUnfolding Techniques
The design and implementation of a temperature control device for the atomic force microscope (AFM) are described. This device is based on a thermoelectric module which can be used for both heating and cooling the samples in the AFM liquid chamber within the range of 5–50°C. A thermocouple is inserted in the liquid chamber to measure the sample temperature. A commercial thermoelectric temperature controller is used to keep the temperature constant during a measurement, which can be stabilized within 0.1°C. To dissipate the heat generated by the thermoelectric module during cooling, a water cooled heat sink is used. Using this device, imaging and mechanical unfolding experiments were carried out at different temperatures. The results show that the temperature control device does not significantly reduce the imaging capacities of the AFM, and that the force-induced unfolding of individual protein molecules can be readily observed at different temperatures. Temperature dependent single molecule measurements can broaden the applications of AFM and reveal new insights into the macromolecular structures and processes.
26
Calibration of atomic-force microscope tips
Jeffrey L. Hutter, John Bechhoefer · Review of Scientific Instruments · 1993 · 4.2K citations
Reversible Unfolding of Individual Titin Immunoglobulin Domains by AFM
Matthias Rief, Mathias Gautel, Filipp Oesterhelt et al. · Science · 1997 · 3.1K citations
Short cantilevers for atomic force microscopy
D. A. Walters, J. P. Cleveland, Neil H. Thomson et al. · Review of Scientific Instruments · 1996 · 490 citations