Measurement Science Review · 2013 · 50 citations · 17 references
EngineeringBioengineeringBiomedical EngineeringBiomechanicsMechanical DesignSurgical InstrumentationContact MechanicsForce SensingForce SensorsMechanical MeasurementMechanical SensorsTactile SensingContact Force EstimationContact Force ModelingCoupling Error AnalysisDecouplingForce Sensing In SurgeryForce-torque SensingRobust DesignTool-tissue Interaction
This paper analyzes the decoupling of a sliding structure six-axis force/torque sensor, which is used to measure the interactive force between surgical tools and soft tissue for the establishment of soft-tissue force model. Because this decoupling structure requires accurate sliding clearance and symmetric grooves, the influence of contact force between the elastic body and the groove sidewall on decoupling is analyzed. The analysis results indicate that the contact force will produce additional coupling error. The robust design method of elastic body size optimization is used to eliminate the influence of contact force. In the calibration test, the expanded uncertainty of the calibration device is evaluated and the calibration results validate the good decoupling.
17
Biomechanics. Mechanical Properties of Living Tissues
Y. C. Fung, Richard Skalak · Journal of Applied Mechanics · 1982 · 6.1K citations · Full text
Guide to the Expression of Uncertainty in Measurement
Barry N. Taylor, C. E. Kuyatt · 1995 · 821 citations