Journal of Physics E Scientific Instruments · 1980 · 274 citations · 10 references
Natural SciencesMaterials EngineeringMaterials ScienceInstrumentationOpticsRadiation MeasurementDetector SystemsDetector TechnologyRadiation DetectorsX-ray Data AnalysisX-ray DiffractionX-ray OpticsCrystallographic Data ReductionDiffraction Data CollectionPowder DiffractionPowder Diffraction AnalysisPowder X-ray DiffractionX-ray CollimatorsX-ray Diffraction Detection
A computer-controlled microdensitometer has been designed and built in this laboratory for measuring Guinier powder diffraction photographs. It consists of a microcomputer-controlled single-beam microdensitometer using as a measuring unit an IR emitting diode ( lambda =940 nm), a collimator system and a photodiode. The scan process is controlled by a microcomputer, Zilog MCZ 1/05, with two floppy disc units. The microcomputer is also used for data reduction. A direct connection with a large scale computer, Amdahl 470/V7, makes it possible to run indexing, search-match or profile refinement programs within 15 minutes from the moment the film is mounted in the scanner.
10
Automatic Densitometer Measurements of Powder Diffraction Photographs.
Gunnar Malmros, Per‐Erik Werner, C. Hebrew et al. · Acta chemica Scandinavica/Acta chemica Scandinavica. B, Organic chemistry and biochemistry/Acta chemica Scandinavica. A, Physical and inorganic chemistry/Acta chemica Scandinavica. Series B. Organic chemistry and biochemistry/Acta chemica Scandinavica. Series A, Physical and inorganic chemistry · 1973 · 52 citations · Full text