Journal of Agricultural and Food Chemistry · 1991 · 46 citations · 0 references
AgricultureFood ScienceNutritionAnalytical ChemistryChromatographic AnalysisFood AnalysisBioanalytical Method ValidationChromatographic Food AnalysisInstrumental AnalysisMineral AnalysisVitamin Analysis
An improved method for the ion chromatographic determination of phytie acid is presented. The method is based on a recently published chromatographic method combined with a treatment of the samples with EDTA. Separation and quantitation of phytie acid were performed on an anion-exchange column. Phytie acid was detected at 300 nm after reaction with a ferric salt in an in-line postcolumn derivatization. The improved method described here was compared with a recently published HPLC method without EDTA treatment and with a method using small single-use ion-exchange columns. It was found that the HPLC method without EDTA treatment of the samples results in values too low for phytie acid, whereas the method using small single-use ion-exchange columns gives values too high. The differences between the results of the three methods were statistically significant (p < 0.01). © 1991 American Chemical Society.