Review of Scientific Instruments · 1996 · 22 citations · 13 references
Natural SciencesElectrical EngineeringMaterials ScienceSemiconductor MaterialsSpectroscopySurface ScienceDevice CharacterizationElectrical CharacterizationReference ElectrodesSemiconductor SurfacesSurface PhysicsSurface SpectroscopyCapacitanceParasitic CapacitanceParasitic Capacitance ReductionReference Electrode DesignSurface Potential MeasurementWork FunctionWork Function DeterminationWork Function Measurement
This paper reports the design of a compact, inexpensive, and ultrahigh vacuum compatible reed-type Kelvin probe for work function and surface photovoltage spectroscopy (SPS) studies. A simple method of controlling the vibration amplitude of the probe is proposed which reduces mechanical instabilities of the vibrating capacitor, leading to an autonomous and reliable probe for monitoring work function changes over a prolonged period of time. A boss-type reference gold electrode is used to simultaneously study the work function changes with adsorbents and SPS on real surfaces of semiconductors. A current preamplifier used in the measuring circuit reduces the effect of parasitic input capacitance and facilitates the application of a biasing voltage. A feedback loop system and off-null technique are used for obtaining the contact potential difference. With this improved design of the probe setup a sensitivity of better than 0.1 mV has been achieved. Some experimental results are presented to show the performance of the setup.
13
Surface Properties of Germanium
W. H. Brattain, John Bardeen · Bell System Technical Journal · 1953 · 590 citations
Iain D. Baikie, Susan MacKenzie, P. J. Estrup et al. · Review of Scientific Instruments · 1991 · 204 citations · Full text