A combined Thomson–Rayleigh scattering diagnostic using an intensified photodiode array

M. C. M. van de Sanden, Guido Janssen, J. M. de Regt, D.C. Schram, J.A.M. van der Mullen, B. van der Sijde

Review of Scientific Instruments · 1992 · 118 citations · 19 references

Concepts

Abstract

A combined Thomson–Rayleigh scattering device is discussed. It consists of a Nd:YAG laser as a light source in combination with a multichannel detection technique consisting of a gated light amplifier in combination with an optical multichannel analyzer. Special attention is focused on the analysis of the measured spectra. Including convolution methods and taking into account weak coherent effects increases the dynamic range and the accuracy of the measured electron density ne and temperature Te and neutral particle density n0. Accuracies of 1%–4% for ne, 2%–6% for Te, and 10%–50% for n0 depending on the plasma condition are obtained. The dynamic range for ne is 7×1017–1021 m−3, for n0 is 1020–1023 m−3 and for Te is 1000–50 000 K.

References

19