Concepedia

Abstract

For the first time, a dosimeter employing two floating gate radiation field effect transistors (FGRADFET) and operating at mere 0.1 V is presented. The novel dosimeter requires no power during irradiation and consumes only 1 during readout. Besides the low power operation, structural changes at the device level have enhanced the sensitivity of the dosimeter considerably as compared to previous designs. The dosimeter is integrated with a wireless transmitter chip, thus eliminating all unwanted communication and power cables. It has been realized monolithically in DALSA's 0.8 complementary metal-oxide-semiconductor process and characterized with X-ray and γ-ray sources. A maximum sensitivity of 5 mV/rad for X-rays and 1.1 mV/rad for -rays have been achieved in measurements. Due to its small size, low-power, and wireless operation, the design is highly suitable for miniaturized, wearable, and battery operated dosimeters intended for radiotherapy and space applications.

References

YearCitations

1986

178

2004

134

2004

76

1998

64

2007

51

2001

50

2008

49

2009

26

1997

26

1996

24

Page 1