International Journal of Occupational Safety and Ergonomics · 2005 · 28 citations · 14 references
PharmacyPharmaceutical AnalysisAnalytical ChemistryChromatographic AnalysisOccupational ToxicologyAnalytical ChromatographyChromatographyInstrumental AnalysisLiquid ChromatographyMobile Phase OptimizationBenzene ExposureConcentrationHigh-performance Liquid ChromatographyHigh-performance Liquid Chromatography Method DevelopmentOccupational Exposure BiomarkersPh-dependent ExtractionSample ConcentrationSample Pretreatment
Factors affecting solid phase extraction (SPE) of trans,trans-muconic acid (ttMA), as a benzene biomarker, including sample pH, sample concentration, sample volume, sample flow rate, washing solvent, elution solvent, and type of sorbent were evaluated. Extracted samples were determined by HPLC-UV (high performance liquid chromatography-ultraviolet). The analytical column was C18, UV wave length was 259 nm, and the mobile phase was H(2)O/methanol/acetic acid run at flow rate of 1 ml/min. A strong anion exchange silica cartridge was found successful in simplifying SPE. There was a significant difference between recoveries of ttMA when different factors were used (p < .001). An optimum recovery was obtained when sample pH was adjusted at 7. There was no significant difference when different sample concentrations were used (p > .05). The optimized method was then validated with 3 different pools of samples showing good reproducibility over 6 consecutive days and 6 within-day experiments.
14
P. Ducos, R. Gaudin, Alain Robert et al. · International Archives of Occupational and Environmental Health · 1990 · 160 citations
Sample preparation for chromatographic separations: an overview
Salwa K. Poole, Thomas A. Dean, John W. Oudsema et al. · Analytica Chimica Acta · 1990 · 147 citations