Analytical Letters · 2008 · 13 citations · 16 references
PharmacyToxicologyAnalytical ChemistryAnalytical ToxicologyChromatographic AnalysisAnalytical ChromatographyBioanalytical Method ValidationChromatographyLiquid ChromatographyAssay DeterminationBioanalytical ChromatographyConcentrationDetection Limit AssessmentMethod Detection LimitOchratoxin A ContaminationOchratoxin ExposureSerum Sampling
Abstract A high‐performance liquid chromatography (HPLC) method has been developed for the determination of ochratoxin A (OTA) in human blood serum. Samples were purified on a C18 solid phase extraction column. The developed method required a relatively low serum volume (0.5 ml). Significant correlation (r of 0.998) was found over the range from 0.10 to 8 ng/ml, with a detection limit of 0.1 ng/ml and better performance in terms of precision and accuracy. Mean recoveries at 0.5 and 2 ng/ml were respectively 69.7±1.2 and 71.9±2.8%. This method was used as a rapid and noninvasive tool to assess human exposure to OTA. Among 40 analyzed serum samples, 27.5% were found to contain OTA with levels going from 0.1 to 11.98 ng/ml with a mean concentration of 0.73±2.35 ng/ml.
16
B. Zimmerli, Rudolf Dick · Journal of Chromatography B Biomedical Sciences and Applications · 1995 · 384 citations
Mycotoxins, endemic nephropathy and urinary tract tumours.
M. Castegnaro · PubMed · 1991 · 188 citations
Medicine, Nephrology, Oncology +14
Human ochratoxicosis and nephropathy in Egypt: A preliminary study
Ehab W. Wafa, Raïda Yahya, Mohamed Sobh et al. · Human & Experimental Toxicology · 1998 · 123 citations
Medicine, Nephrology, Pathology +16