Characterization of a Disposable Electrochemical Biosensor Based on Putrescine Oxidase from <i>Micrococcus rubens</i> for the Determination of Putrescine

Wilder Henao-Escobar, Olga Domínguez‐Renedo, M. Asunción Alonso‐Lomillo, José F. Cascalheira, A.C. Dias-Cabral, M. Julia Arcos‐Martínez

Electroanalysis · 2014 · 10 citations · 37 references

Concepts

Abstract

Abstract Putrescine oxidase (PUO) from Micrococcus rubens has been characterized, as a biocatalytic component for the development of electrochemical biosensors for the determination of putrescine. PUO was immobilized by cross‐linking on screen‐printed carbon electrodes (SPCEs) containing tetrathiafulvalene (TTF) as mediator (TTF/SPCEs). Isothermal titration calorimetry was used for investigation of kinetic properties of immobilized PUO. These PUO/TTF/SPCEs showed a linear response range from 10 a 74 µM, a capability of detection of 10.1±0.6 µM ( n =4, α = β =0.05) and a precision of 6.7 %. Octopus and zucchini samples have been analysed using HPLC as a reference technique.

References

37