Natural SciencesMaterials EngineeringMaterials ScienceAnalytical ChemistryElectronic MaterialsSpectroscopyElemental ImpuritiesX-ray Data AnalysisX-ray DetectionX-ray DetectorsX-ray Fluorescence AnalysisX-ray SpectroscopyImpuritiesSilicon WafersSolid-state X-ray DetectorsTotal Reflection X-ray FluorescenceTrace Element ImpuritiesWafer MaterialsX-ray Fluorescence SpectroscopyX-ray Reflection
Spurious peaks such as Fe-Kα or Ni-Kα radiations are observed in the analysis of trace metallic impurities on a Si wafer by monochromatic total reflection X-ray fluorescence. These peaks seriously influence the accuracy in the quantitative analysis of the trace impurities. Intensities of these peaks are shown to depend on the intensity of the elastically scattered primary beam ( W-Lβ) whose intensity varies with the incident azimuth. It is found that impurity radiations are included in the primary X-ray beam, and contaminations in the Be window of the X-ray detector are also a source of spurious peaks. The impurities deposited on the Be window, in and/or on the Si(Li) crystal of the detector and the casing material of the X-ray detector can also contribute to these spurious peaks.