Journal of Medical Devices · 2009 · 84 citations · 71 references
EngineeringRehabilitation MedicineRobotics EngineeringRehabilitationRehabilitation EngineeringRoboticsActuation SystemsActuator IntegrationCoordination TrainingRehabilitation RoboticsRobotic ExoskeletonsRobotic RehabilitationBrakesExoskeleton EvaluationExoskeleton RehabilitationExoskeletonsForce CoordinationHydraulic BrakesPassive ExoskeletonsRotations
The Dampace exoskeleton combines functional exercises resembling activities of daily living with impairment-targeted force-coordination training. The goal of this paper is to evaluate the performance of the Dampace. In the design, the joint rotations are decoupled from the joint translations; the robot axes align themselves to the anatomical axes, overcoming some of the traditional difficulties of exoskeletons. Setup times are reduced to mere minutes and static reaction forces are kept to a minimum. The Dampace uses hydraulic disk brakes, which can resist rotations with up to 50 N m and have a torque bandwidth of 10 Hz for multisine torques of 20 N m. The brakes provide passive control over the movement; the patients’ movements can be selectively resisted, but active movement assistance is impossible and virtual environments are restricted. However, passive actuators are inherently safe and force active patient participation. In conclusion, the Dampace is well suited to offer force-coordination training with functional exercises.
71
Gill A. Pratt, Matthew M. Williamson · 2002 · 2.2K citations
Engineering, Mechanical Engineering, Robotics Engineering +17
Design and Evaluation of the LOPES Exoskeleton Robot for Interactive Gait Rehabilitation
Jan F. Veneman, R. Kruidhof, Edsko E.G. Hekman et al. · IEEE Transactions on Neural Systems and Rehabilitation Engineering · 2007 · 1.3K citations · Full text
Engineering, Rehabilitation Medicine, Robotics Engineering +17